Ii.2 spindle selection, Ii.2, Spindle selection – Brookfield DV-I Viscometer User Manual
Page 10: Sss s s
Brookfield Engineering Laboratories, Inc.
Page 10
Manual No. M/92-021-O604
cP 123E3 70.1°F
0.0RPM % 0.0
cP 0.0 S01
0.0RPM % 0.0
(with Temperature Probe)
(without Temperature Probe)
Calculated Viscosity (cP or m•Pas)
Speed
Temperature (if available)
% Torque
Spindle Selected
Figure II-5
The display will vary slightly depending upon the status of the last spindle entry.
NOTE:
If the viscosity value exceeds 99,999, scientific notation is used. In FigureII-5, the
viscosity value is 123,000 cP.
II.2 Spindle Selection
LVDV-I+ Viscometers are provided with a set of four spindles and a narrow guardleg; RVDV-I+
Viscometers come with a set of six spindles and a wider guardleg; HADV-I+ and HBDV-I+
Viscometers come with a set of six spindles and no guardleg. (See Appendix E for more informa-
tion on the guardleg.)
The spindles are attached to the viscometer by screwing them on to the lower shaft. Note that the
spindles have a left-hand thread. The lower shaft should be held in one hand and lifted up. The
spindles should be screwed to the left. The face of the spindle nut and the matching surface on the
lower shaft should be smooth and clean to prevent eccentric rotation of the spindle. Spindles can
be identified by the number on the side of the spindle nut.
The DV-I+ requires a Spindle Entry Code number to calculate viscosity values. The two digit
entry code for each spindle may be found in Appendix D.
NOTE:
The DV-I+ will remember the Spindle Entry Code which was in use when power
was turned off.
II.2.1 Spindle Selection for Models WITHOUT Temperature Display
Pressing the
SELECT SPINDLE
key will cause the characters on the top line of the display to
begin to blink . It will blink for about three seconds. If the
UP
or
DOWN
ARROW
keys are pressed
(while
S
SS
S
S
is blinking) the two character spindle value to the right of the
S
SS
S
S
character will begin to
change (in either an increasing or decreasing direction depending upon which
ARROW
key is
pressed) for each press of the key. If the
ARROW
key is pressed and held, the display will scroll
through the spindle codes for as long as the
ARROW
key is depressed. When it reaches the last
item in the list (either at the top or bottom of the list) the spindle code displayed will “roll-over”
to either the first or last spindle code and the scroll action will continue.
When the desired spindle code is displayed, release the
ARROW
key to halt further scrolling.
Press the
SELECT SPINDLE
key once again. This will cause the
S
SS
S
S
character to cease blinking and
the new spindle code will be accepted for use in viscometer calculations.