Appendix E - Calibration Procedures

The accuracy of the DV-I+ is verified using viscosity standard fluids which are available from Brookfield
Engineering Laboratories or your local Brookfield agent. Viscosity standards are Newtonian, and
therefore, have the same viscosity regardless of spindle speed (or shear rate). Viscosity standards,
calibrated at 25

°

C, are shown in Table E-1 (Silicone Oils) and Table E-2 (Mineral Oils).

Container size:

For Viscosity Standards < 30,000 cP, use a 600 ml Low Form Griffin
Beaker having a working volume of 500 ml.

For Viscosity Standards

30,000 cP, use the fluid container.

Inside Diameter: 3.25"(8.25cm)
Height:

4.75"(12.1cm)

Note: Container may be larger, but may not be smaller.

Temperature:

As stated on the fluid standard label: (

+

/-) 0.1

°

C

Conditions:

The DV-I

+

should be set according to the operating instructions. The water

bath must be stabilized at test temperature. Viscometers with the letters “LV
or “RV” in the model designation must have the guard leg attached (see page
34 for more information on the guard leg).

Normal 25

°

C Standard Fluids

High Temperature Standard Fluids

Viscosity (cP)

Viscosity (cP)

Three Viscosity/Temperatures**

5

5,000

HT-30,000

10

12,500

HT-60,000

50

30,000

HT-100,000

100

60,000

500

100,000

**25

°

C, 93.3

°

C, 149

°

C

1,000

Refer to Brookfield catalog for

more information.

Table E-1

MINERAL OIL VISCOSITY STANDARD FLUIDS

BEL Part No.

Viscosity (cP) 25

°

C

B31

31

B210

210

B750

750

B1400

1,400

B2000

2,000

B11000

11,000

B20000

20,000

B80000

80,000

B200000

200,000

B420000

420,000

Table E-2