beautypg.com

Appendix b - viscosity ranges, Appendix b, Viscosity ranges – Brookfield DV-I Viscometer User Manual

Page 24: Lv(#1-4) and rv,ha,hb(#1-7) viscometers, Vane spindles

background image

Brookfield Engineering Laboratories, Inc.

Page 24

Manual No. M/92-021-O604

Appendix B - Viscosity Ranges

LV(#1-4) and RV,HA,HB(#1-7) Viscometers

Viscometer

Minimum

Maximum

LVDV-I+

15

2,000,000

RVDV-I+

100

13,300,000

HADV-I+

200

26,600,000

HBDV-I+

800

106,400,000

Viscosity Range (cP)

Vane Spindles

V-71

RV

.5-5

5-50

262 -

2620

V-72

RV

2-20

20-200

1110 -

11100

V-73

RV

10-100

100-1000

5350 -

53500

V-74

(4)

V-71

HA

1-10

10-100

524 -

5240

V-72

HA

4-40

40-400

2220 -

22200

V-73

HA

20-200

200-2000

10700 -

107000

V-74

(4)

V-71

HB

4-40

40-400

2096 -

20960

V-72

HB

16-160

160-1600

8880 -

88800

V-73

HB

80-800

800-8000

42800 -

428000

V-74

(4)

V-71

5xHB

20-200

200-2000

10480 -

104800

V-72

5xHB

80-800

800-8000

44400 -

444000

V-73

5xHB

400-4000

4000-40000

214000 - 2140000

V-74

(4)

Spindle

Torque

Shear Stress Range

Viscosity Range cP

Range

Pa

dyne/cm

(mPa•s) @ 10 rpm

Notes:

1) 1 Pa = 10 dyne/cm
2) 1 cP = 1 mPa•s
3) Possibility of turbulence at speeds above 10 rpm may give artificially higher viscosity readings.

2

2

(4)

V-74 not in the firmware for the DV-I+ spindle table. Manual calculations are required to convert
torque readings to shear stress and viscosity values.