Accuracies and ranges, 2 accuracies and ranges – Eclipse Combustion EGA4 User Manual
Page 7
![background image](/manuals/65249/7/background.png)
Instruction Manual MM850481 ed.01
1.2
Accuracies and ranges
Parameter
Sensor type
Range
Resol.
Max
response
Accuracy
O
2
Electrochemical
from 0 to 25.0%
0.1%
20 sec.
±0.1% vol.
CO
H
2
compensated up
to 1000ppm
Electrochemical
from 0 to 8000 ppm
1 ppm
50 sec.
±10 ppm <300 ppm
±4% rdg up to 2000 ppm
±10% rdg >2000 ppm
CO
Electrochemical
from 0 to 20000 ppm
1 ppm
40 sec.
±10 ppm <300 ppm
±4% rdg up to 2000 ppm
±10% rdg >2000 ppm
NO
Electrochemical
from 0 to 4000 ppm
1 ppm
40 sec.
±5 ppm <125 ppm
±4% rdg >125 ppm
NO
2
Electrochemical
from 0 to 1000 ppm
1 ppm
50 sec.
±5 ppm <125 ppm
±4% rdg >125 ppm
NO
x
Calculated
from 0 to 5000 ppm
1 ppm
SO
2
Electrochemical
from 0 to 4000 ppm
1 ppm
40 sec.
±5 ppm <125 ppm
±4% rdg >125 ppm
CO
2
Calculated
from 0 to 100%
0.1 %
Tair
Pt100
from -10 to 100 °C
0.1 °C
±(0.2% rdg+0.15°C)
Tgas
Tc K
from -10 to 1000 °C
0.1 °C
±(0.3% rdg+0.3°C)
∆T
Calculated
from 0 to 1000 °C
0.1 °C
Tflow
Tc K
from -10 to 100 °C
0.1 °C
±(0.3%rdg+0.3°C)
Treturn
Tc K
from -10 to 100 °C
0.1 °C
±(0.3% rdg+0.3°C)
Pressure/Draft &
∆P
Piezo
-10hPa to +100 hPa
0.01 hPa
±1% rdg
Excess air
Calculated
from 1.00 to infinite
0.01
Efficiency
Calculated
from 0 to 100%
0.1 %
Gas speed
Pitot tube
From 0 to 100.0 m/s
0.1 m/s
Smoke index
External pump
from 0 to 9
Technical units and ranges can be converted directly from ppm to mg/Nm
3
, mg/kwh and from hPa to mmH
2
O, mbar or inH
2
O.
The relative accuracy shown are expressed as absolute or % of rdg errors at -5°C to +40°C ambient temperature.
The maximum response time shown is referred to 90% signal changes.
The pressure relative accuracy shown is valid only after the auto-zero procedure.
Specifications may change without notice.
7