Maintenance, Error messages, Set-up menu – Eclipse Combustion EGA4 User Manual
Page 43: 10 maintenance, 1 error messages, 2 set-up menu
Instruction Manual MM850481 ed.01
10 MAINTENANCE
The EGA4 analyzer has been developed and manufactured using high quality components. A correct, systematic
maintenance will prevent damage and will increase substantially the life of the instrument, giving excellent results.
Anomalous conditions that require extraordinary maintenance interventions are promptly displayed, either clearly or with
an information code.
10.1 Error
messages
The instrument performs the autodiagnostic procedure when switched on. When measuring it checks also for
anomalous conditions. If analyzer founds trouble, it display a message on LCD. Sometime these messages corresponds
to a unrecoverable problem and it is necessary to return the instrument to Eclipse for maintenance. Frequently these
message are referred to a recoverable trouble and can be solved by yourself.
The following table show the possible instrument error messages:
Error message
Problem
Solutions
Calc Err
Parameter calculus error.
The instrument cannot perform the
calculus. Check and solve the others error
conditions.
Probe Er
The external Rtd probe is
unconnected or damaged.
Connect or replace the external
temperature probe.
RJ Err
Internal temperature reference
damaged.
Return the instrument to Eclipse for
maintenance
Over
Overrange condition
The problem is solved automatically when
measure return inside the range
Under
Underrange condition
The problem is solved automatically when
measure return inside the range
--------
No calculus available
The instrument cannot perform the
calculus. Check and solve the others error
conditions. (e.g. O
2
>18%).
Excluded
CO sensor excluded by electro-valve Automatic CO sensor protection is active.
Perform one or more autozero cycle in
fresh air.
Tc Error
The thermocouple is unconnected or
damaged.
Connect or replace the temperature probe.
Tsens Er
Cell compensation error
Return the instrument to Eclipse for
maintenance.
Zero Err
The autozero procedure is failed
Perform one or more autozero cycle in
fresh air. If the problem remain, Return the
instrument to Eclipse for maintenance.
Sens. Er
Broken
cell
Return the instrument to Eclipse for
maintenance.
No meas.
Empty Memory
No sens.
The cell is non installed
No probe
No external probe connected on
mini-din connector.
Connect or replace the temperature probe.
10.2 Set-up
menu
Configuration menu is protected by a password. This procedure menu allows the operator to change some internal
parameters values: Date and time internal clock, communication baud rate, autozero time and leak test timing.
Follow the procedure below to activate the procedure:
Switch the instrument off
43