10 MAINTENANCE
The EGA4 analyzer has been developed and manufactured using high quality components. A correct, systematic 
maintenance will prevent damage and will increase substantially the life of the instrument, giving excellent results. 
Anomalous conditions that require extraordinary maintenance interventions are promptly displayed, either clearly or with 
an information code. 
10.1 Error
messages
The instrument performs the autodiagnostic procedure when switched on. When measuring it checks also for 
anomalous conditions. If analyzer founds trouble, it display a message on LCD. Sometime these messages corresponds 
to a unrecoverable problem and it is necessary to return the instrument to Eclipse for maintenance. Frequently these 
message are referred to a recoverable trouble and can be solved by yourself. 
The following table show the possible instrument  error messages: 
 
Error message
Problem
Solutions
Calc Err
Parameter calculus error.
The instrument cannot perform the 
calculus. Check and solve the others error 
conditions. 
Probe Er
The external Rtd probe is 
unconnected or damaged. 
Connect or replace the external 
temperature probe. 
RJ Err
Internal temperature reference 
damaged. 
Return the instrument to Eclipse for 
maintenance 
Over
Overrange condition
The problem is solved automatically when 
measure return inside the range 
Under
Underrange condition
The problem is solved automatically when 
measure return inside the range 
--------
No calculus available
The instrument cannot perform the 
calculus. Check and solve the others error 
conditions. (e.g. O
2
>18%).
Excluded
CO sensor excluded by electro-valve Automatic CO sensor protection is active.
Perform one or more autozero cycle in 
fresh air. 
Tc Error
The thermocouple is unconnected or 
damaged. 
Connect or replace the temperature probe.
Tsens Er
Cell compensation error
Return the instrument to Eclipse for 
maintenance. 
Zero Err
 
The autozero procedure is failed 
Perform one or more autozero cycle in 
fresh air. If the problem remain, Return the 
instrument to Eclipse for maintenance. 
Sens. Er
Broken
cell
Return the instrument to Eclipse for 
maintenance. 
No meas.
Empty Memory
No sens.
The cell is non installed
No probe
No external probe connected on 
mini-din connector. 
Connect or replace the temperature probe.
10.2 Set-up
menu
Configuration menu is protected by a password. This procedure menu allows the operator to change some internal 
parameters values: Date and time internal clock, communication baud rate, autozero time and leak test timing. 
 
Follow the procedure below to activate the procedure: 
Switch the instrument off
43