Flaps calibration, Flaps calibration -7 – Dynon Avionics EMS-D10 Installation Guide User Manual
Page 49
EMS
Configuration
pressing NEXT. Repeat the process for the opposite position. The process will then prompt you
to put the trim into takeoff position. If you do not require a takeoff indication on the given axis’
trim display, you may press NONE. When you have completed the calibration, press the DONE
button.
Press the TAKOFF button to calibrate the takeoff position indicator. When calibrated, a green
line is displayed on the trim scales, indicating takeoff position.
View the trim display on the EMS Main or Aux page (depending on where you configured it to
display) to make sure that it works as expected. You may repeat this calibration process as many
times as you wish.
Flaps Calibration
EMS-D10 Installation Guide
5-7
Prior to calibrating your flaps sensor, ensure that you have connected it as described
on page 3-15 and selected FLAP POS for the desired GP input FUNCT parameter
as described on page 5-17.
Enter the EMS menu by pressing any button below an EMS main page and pressing
MORE > SETUP > MORE > FLAPS. The FLAP CALIBRATION menu just s
FLAPS, followed by the GP input that each is configured as (or NONE, if no GP
input is configured for a given axis). Ensure that this list corresponds to the phys
connections made during the setup described in the
hows
ical
General Purpose Inputs section
on page 3-13. Press SEL► to enter the flaps calibration menu.
During the calibration process, ensure that the number shown in the VALUE field changes as
you adjust the flaps. If the number does not change, the flaps sensor may be incorrectly wired to
the EMS-D10, or incorrectly configured in the GP input selection under the SENSOR menu. At
any point in the process, you may press CANCEL to end the calibration without overwriting the
previous calibration results.
You must calibrate for at least 2 positions, and may calibrate for as many as 5. The calibration
process first requires you to put the flaps in the 0° extended position. When you have done this,
press NEXT. Next, you will calibrate for the second position. Press INC+ or DEC- to set the
angle that you would like displayed for the second position. If you only wish to have 2 positions
displayed, press DONE. Otherwise, press NEXT to repeat the process for the third position.
When you have completed the calibration, press DONE. View the flaps display on the EMS
Main or Aux page (depending on where you configured it to display) to make sure that it works
as expected. You may repeat this calibration process as many times as you wish.