Fi-90 specifications – Dillon FI-90 User Manual
Page 4
4
FI-90 Specifications
Display:
7-segment LCD, 6-1/2 digits plus sign, .5 inch high, movable decimal, with
7 annunciators
Display Update Rate:
Two times per second
Analog-to-Digital
Conversion Rate:
60 times per second
Resolution:
Up to 10,000 divisions standard
Controls:
Pushbutton zero; pushbutton peak-hold; pound-, kilogram-, Newton-scroll selection;
PRINT, MENU, and SELECT keys for configuration; ON / OFF switch (optional: used
only on alkaline battery option); BATT / AC switch (optional: used on rechargeable
battery / AC versions)
Peak Detection Rate:
60 times per second
Peak Detection:
Threshold 1% of capacity setting
Accuracy :
Span: ±5.0 ppm/C
Zero: ±.066 uV/C (-10 to 40°C)
Linearity:
±0.005% of capacity, maximum
Repeatability:
±0.005% of capacity, maximum
Hysteresis:
0.005% of capacity, maximum
Power:
Std - 117VAC +10/-15% 50/60Hz
5 watt
Opt - 230VAC +10/-15% 50/60Hz
5 watt
Opt - 12VDC +10/-15%
130mA max.
Load Cell Drive Capacity:
Up to four 350-ohm load cells. Up to twelve 1000-ohm load cells.
(+5VDC load cell excitation voltage)
Environment:
-10 to 40°C (14 to 104°F)
10 to 90% relative humidity
Calibration
and Programming:
All calibration and programming is done through the front panel with data stored in
non-volatile memory.
Analog Range:
-0.14 to +3.5 mV/V
Scale Capacities:
.00001 to 999999, programmable to any number between these limits
Scale Division Sizes:
.0001 to 20000, programmable to any division size between these limits
Pushbutton-Zero Range:
0 to ±100% of capacity; programmable in any percentage of indicator capacity;
independent positive and negative limits; unit will not allow zeroing beyond capacity.
Overrange Capacity:
The indicator displays force measurements up to and including full-scale capacity