Welch Allyn Suretemp Plus Calibration/Verification Module Notes - Quick Reference Guide User Manual
Page 6

Speci
es
Code Definition / Description Origin/Alert
type
Traditional
code map
(ST+ OEM)
Traditio
nal code
map
(ST+
HH)
Details
Host
controller
action
8
9
Missing probe well.
Probe (well)/
Equipment alert
63
E4.9
Insert probe
well.
Indicate
missing
probe well
8
10 Problem saving to the
module's EEPROM in
biotech mode.
Module/ Host
controller alert
65
E4.3
Verify that
batteries are
fresh. Try to
make changes
again. If the
problem
persists,
replace the
module.
Indicate
internal
error
8
11 The module's error
detection mechanism
detected an error.
Module/
Equipment alert
66
E0.8
Internal
malfunction.
Contact Welch
Allyn
Technical
Service.
Indicate
module
error
8
12 The module's probe
error detection
mechanism detected an
error.
Probe/ Equipment
alert
66
E0.8
Probe error,
try again. If
the problem
persists,
replace the
probe.
Indicate
probe
error
8
13 The module's log error
detection mechanism
detected an error.
Module/
Equipment alert
Internal error
log problem;
error log will
be cleared.
Restore
default
error log
8
14 The module's calibration
error detection
mechanism detected an
error.
Module/
Equipment alert
Try again. If
the problem
persists,
replace the
module.
Indicate
module
error
8
15 No probe connected.
Probe/ Equipment
alert
Connect a
probe and try
again. If a
probe is
connected,
replace the
probe. If the
problem
persists
replace the
module.
Indicate
probe
error