Welch Allyn Suretemp Plus Calibration/Verification Module Notes - Quick Reference Guide User Manual
Page 5

Speci
es
Code Definition / Description Origin/Alert
type
Traditional
code map
(ST+ OEM)
Traditio
nal code
map
(ST+
HH)
Details
Host
controller
action
F),
respectively.
If conditions
are valid and
the problem
persists,
replace the
probe. If
problem
persists,
replace
module.
8
3
Internal calibration
resistor (RCAL) on the
board is damaged or
contaminated (pulse too
long).
Module/
Equipment alert
47
E4.1
Internal
malfunction.
Contact Welch
Allyn
Technical
Service.
Indicate
module
error
8
4
Internal calibration
resistor (RCAL) on the
board is damaged or
contaminated (pulse too
short).
Module/
Equipment alert
48
E4.1
Internal
malfunction.
Contact Welch
Allyn
Technical
Service.
Indicate
module
error
8
5
Internal circuit
validation resistor (PTB)
on the board is damaged
(value is over).
Module/
Equipment alert
49
E4.0
Internal
malfunction.
Contact Welch
Allyn
Technical
Service.
Indicate
module
error
8
6
Internal circuit
validation resistor (PTB)
on the board is damaged
(value is under).
Module/
Equipment alert
50
E4.0
Internal
malfunction.
Contact Welch
Allyn
Technical
Service.
Indicate
module
error
8
7
A/D measurement timed
out.
Module/
Equipment alert
51
E0.2
Internal
malfunction.
Contact Welch
Allyn
Technical
Service.
Indicate
module
error
8
8
The probe was not
characterized/calibrated.
Probe/ Equipment
alert
52
E0.9
Probe
malfunction;
replace probe.
Indicate
probe
error