Suretemp plus error codes – Welch Allyn Suretemp Plus Calibration/Verification Module Notes - Quick Reference Guide User Manual
Page 2

SureTemp Plus error codes
To find the SureTemp Plus error code:
1. Go to the tab of the device you are working on and select the SureTemp Plus module.
2. Select the View log button in the user information section of the page. A .txt document opens with the last 10
SureTemp Plus error codes. The last listed error code is the latest error code.
The module provides these errors in a CErrEStd object. For a unique error code, merge the identified species value
with the error code. If the status indicates a fault condition, the module suggests recovery attempts, which may
include the host initializing a reset.
The errors are grouped into the following origins:
• Probe
• Module
• Ambient
• Power
• Patient
Errors are also classified as the following alert types:
• Host controller
• Equipment
• User
Speci
es
Code Definition / Description Origin/Alert
type
Traditional
code map
(ST+ OEM)
Traditio
nal code
map
(ST+
HH)
Details
Host
controller
action
0
0
No error - used for ACK
messages in response to
commands.
Module/ Host
controller alert
Internal error
None
None
1
5
WACP message CRC
mismatch.
Module/ Host
controller alert
Internal error
None
None
2
1
This message is not
implemented by the
module.
Module/ Host
controller alert
Internal error
None
None
2
2
This message is not
supported by the
module.
Module/ Host
controller alert
Internal error
None
None
2
3
The module has run out
of memory.
Module/ Host
controller alert
Internal error
None
None
2
5
The parameter provided
is invalid for the
specified message.
Module/ Host
controller alert
Internal error
None
None
2
6
The parameter provided
is outside of the
allowable range for the
specified message.
Module/ Host
controller alert
Internal error
None
None