Performing an Auto ECG test

CAUTION Patient data is not saved until the ECG test is completed.

Note

The ECG configuration settings can be changed in the Settings tab. The
following settings may appear differently if the default settings have been
modified.

1. Touch the

(Auto ECG) button. The Summary tab appears.

2. Enter the following patient information as desired:

Patient ID. Touch the OK button.

Birth date. Touch the OK button.

Last name. Touch the OK button.

First name. Touch the OK button.

Middle Initial. Touch the OK button.

Note

If the patient has a pacemaker touch the Pacemaker present
button.

3. Touch the (Next) button.
4. Enter the following patient information as desired:

Gender

Race

Height. Touch the OK button.

Weight. Touch the OK button.

Comments. Touch the OK button.

5. Touch the (Next) button.
6. Attach the leads to the patient.

Note

If desired, touch the

(torso) button to enlarge the

lead-placement screen. Any flashing dots on the ECG
preview screen indicate unattached or poorly attached leads.

7