Welch Allyn CP150 12-Lead Resting Electrocardiograph, Software Version 2.00.XX - Quick Reference Guide User Manual
Page 10

Item
Button
1
Leads button
2
ECG preview button
3
Gain button (size)
4
Speed button
5
Filters button
7. If an Artifact message appears, minimize the artifact, as described under
Troubleshooting. You might need to ensure that the patient is comfortably warm,
reprepare the patient’s skin, use fresh electrodes, or minimize patient motion.
8. (Optional) Adjust the waveforms, using the buttons to cycle through the following
options:
•
leads displayed
•
ECG preview format
•
gain (size)
•
speed
•
filters
9. Touch the Print button to perform the Auto ECG test.
If Print preview on is selected in the ECG configuration settings, a single screen of
the ECG Report Print Preview appears. Touch Print to continue with the Auto ECG
test or touch Retest to discard the test and return to the previous screen.
10. If a Waiting for 10 seconds of quality data message appears, at least 10 seconds of
ECG data have been collected with excessive artifact. Time requirements in the
message may vary, based upon selected print format. Minimize the artifact, as
described under Troubleshooting. Then wait for the test to print. If necessary, you
can override the wait time and print the available data immediately, but the result
might be an incomplete or poor-quality test.
8 Performing an Auto ECG test
CP 150 12-lead resting electrocardiograph