beautypg.com

Specification changes, Clarification of propaq encore specification, Propaq 100 series specification changes – Welch Allyn Propaq 6.2X-8.4X/Propaq Encore 1.20-1.3X - Quick Reference Guide User Manual

Page 12

background image

10

Protocol Systems, Inc.

Specification Changes

Clarification of Propaq Encore Specification

The temperature probe tolerance has been restated:

Propaq 100 Series Specification Changes

The following specifications have changed:

Software

Version

Characteristic

Specification

1.20-1.3X

Temperature
Channel
Accuracy

Temperature Range

Tolerance

0 to +10°C

±0.2°C

a

>10 to +50°C

±0.1°C

a

+32 to +50°F

±0.4°F

a

>50 to +122°F

±0.2°F

a

a. Plus probe tolerance.

Software

Version

Characteristic

Specification

8.0X-8.4X

INCO

2

Display

Ranges

8

a

-25 mmHg, 1.1

a

-5kPa, 1.1

a

-5%

a. Lower if in alarm.

8.4X

SpO

2

Range

1% to 100%