Care and cleaning, Trouble-shooting – Welch Allyn PRO 4000 Ear Temperatures with the Braun ThermoScan - User Manual User Manual

Page 6

background image

If the thermometer shows «RET» (without the battery symbol) on its display, the

anti-theft time has elapsed and the thermometer must be returned to the station

for unlocking.

If the anti-theft function is not enabled, the display will show «off» when in stand-

by mode.

Care and cleaning

Braun ThermoScan probe covers are intended for single use only!

To ensure accurate readings, always use Braun ThermoScan disposable probe

covers with the PRO 4000 thermometer. Damaged, perforated, or soiled probe

covers can lead to inaccurate temperature readings. Additionally, the accuracy of

your temperature readings can be affected by damage to the probe window, or

the presence of dirt or cerumen on the probe window.

To assure a high level of accuracy, it is very important to keep this window clean,
dry and undamaged. Fingerprints, cerumen, dust and other soiling compounds

reduce transparency of the window and result in lower temperature measure-

ments. If the thermometer is accidentally placed in the ear without a probe cover

installed, the window must be cleaned immediately. Additionally, the window

and/or probe tip area should never be modified, changed, or adjusted. These

changes will affect calibration and accuracy of the thermometer.

To clean the window, gently wipe its surface with a cotton swab slightly moist-

ened with isopropyl alcohol and immediately wipe dry with a clean cotton swab.

After cleaning, allow at least 5 minutes drying time before taking temperatures.

Make sure probe window is clean and dry. Avoid touching the window except

when cleaning is required. If the window is damaged, the thermometer must be

returned to Welch Allyn for service.

Note: Do not use any chemical other than isopropyl alcohol to clean the probe

window.

Thermometer

Use a soft cloth slightly moistened with isopropyl alcohol to clean the thermome-

ter display and exterior. As needed, clean the thermometer and probe with a 70%

isopropyl alcohol solution or a 10% chlorine bleach solution or a nonstaining

disinfectant such as CaviCide

®

or its equivalent. Do not use abrasive cleaners.

Never submerge the thermometer in water or any other liquid. Store thermometer

and probe covers in a dry location (the thermometer is not protected against in-

gress of water), free from dust and contamination and away from direct sunlight.

The ambient temperature at the storage location should remain fairly constant

and within the range of 50 °F to 104 °F (10 °C to 40 °C).

Trouble-shooting

Situation

Solution

Error message

No probe cover is attached.

Attach new, clean probe cover.

(POS = position error)
The infrared monitor cannot find a

temperature equilibrium and allows

no measurement.

Clear the display by changing

the probe cover.
Ensure that the positioning of the

probe is correct and remains stable.

(flashing)

Ambient temperature is not within the

allowed operating range 50–104 °F

(10–40 °C) or changing too rapidly.

Allow the thermometer to remain for 30

minutes in a room where the tempera-

ture is between 50 and 104 °F / 10 and

40 °C.

Temperature taken is not within typical

human temperature range.

HI will be displayed when temperature

is higher than 108 °F (42.2 °C)

LO will be displayed when temperature

is lower than 68 °F (20 °C)

Make sure a new, clean probe cover is

attached and thermometer is properly

inserted. Then, take a new temperature

System error

If error persists,

If error still persists,

Wait 1 minute until the thermometer

turns off automatically, then turn on

again.

... reset the thermometer by removing

the batteries and putting them back

in.

... contact Welch Allyn.

10

11