48

Chapter 5 Performing ECG Tests

Welch Allyn CP 100 Electrocardiograph

Recording an Auto ECG

An Auto ECG is a report typically showing a 10-second acquisition of 12 leads of ECG
information combined with patient data, interpretation, and measurements matrix. Two
user-defined formats are available: Auto Report 1 or Auto Report 2. To learn how to set up
the Auto ECG report format, see “About Auto ECG Reports” in the Directions for Use.

As shown in the following diagram, there are the two types of Auto ECG: normal and stat.
For details, see these procedures:

“Recording a Normal Auto ECG”

on page 49

“Recording a Stat Auto ECG”

on page 53

Figure 31. Auto ECG Testing, Process Diagram

WARNING To avoid the risk of associating ECG reports with the wrong patients,
each report must identify the patient. If any ECG report does not identify the
patient, write the patient identification patient information on the report
immediately following the ECG test.

Normal

Stat

Press and hold.

Press Quickly.

Enter Patient Data.

Auto ECG Post-Print

screen.

Auto Report prints.

Press Exit