Hand Strap Replacement
1)
Position the unit so that the hand strap is facing up and the battery door com-
partment is toward you.
2)
Push the hand strap forward until the square notches beneath it are visible;
hold the strap in this position with one hand.
3)
Insert the tool provided into one of the notches and press outward until the
fastener is released.
4)
Repeat step 3 on the other fastener.
5)
Insert the fasteners of the new strap into the appropriate slots, so that
“Welch Allyn” can be read.
6)
Finally, pull on the strap to ensure that it is seated properly.
Cleaning and Sterilization of the SureSight
NOTE: Excess solution entering the assembly could damage internal
components. Use caution to ensure cloth is not saturated with solution.
Cleaning solutions should be sprayed on the soft cloth rather than directly
on the SureSight device.
Do not sterilize SureSight or any of its components.
All windows on the SureSight can be cleaned with a soft cloth moistened with
70% Isopropyl Alcohol or any standard window cleaner.
Keep front window clean to insure proper function.
The remaining housings, softgrip, handstrap and stand may be cleaned with
either a dry, soft cloth or a soft cloth moistened with any of the following: 10%
Clorox solution, soapy water, 70% Isopropyl Alcohol, alcohol pads, standard
window cleaner.