Cleaning

Endoscopic instruments should be cleaned immediately after each use.
Endoscopes are delicate and will degrade if not cleaned promptly, due to
the effects of blood, mucous, etc. The steps outlined on the following pages
have been tested and verified to have no damaging effects. Therefore, these
procedures should be adhered to.

Supplies Needed For Cleaning

Large basin of water

Disposable gloves

Cleaning solution (soap solution)

Leakage tester

Gauze pads

Soft scrub brush

Immediately After Procedure…

1. Immediately after removing the instrument from the patient, gently

wipe all blood, mucous and debris from the insertion tube with a
gauze pad moistened in clean water.

2. Turn off the “power”

switch on the light
source.

3. Disconnect endoscope

from light source by
grasping endoscope con-
nector terminal/adaptor
and pulling back gently.