Prior to initial use, System set up – Welch Allyn RL-150 Rhinolaryngoscope - User Manual User Manual
Page 5
Prior to Initial Use
Before set up and inspection of the equipment, check all components
received against the list of components (see Components section, pg. 1) to
verify complete set. If parts are missing, please notify Welch Allyn. Review
the Nomenclature, Set Up, Operation and Cleaning/Disinfection sections to
become familiar with the equipment.
Specifically Inspect:
Rhinolaryngoscope
Insertion Tube—For tears, cuts, dents, bubbles, bumps
Control Section—Test angulation lever/bending section deflection to
assure smooth articulation
Light Source
Cabinet—For any dents, scratches or other surface abnormalities
System Set Up
Rhinolaryngoscope
The No. 65015 adaptor is coupled to your RL-150 Rhinolaryngoscope
endoscope connector terminal when shipped from Welch Allyn. [In the
event that the adaptor should become uncoupled, Steps 1-2 (below)
should be undertaken. Otherwise, proceed to Step 3.]
1. Place the No. 65015 adaptor over the light guide terminal.
2. Thread the adaptor on until snug against endoscope connector terminal.
3. Insert the male end of the adaptor into the Light Source endoscope
connector port. Push until the ball detent in the light source “clicks”
into place to ensure a secure fit.
3
Endoscope
Connector
Terminal
RL-150 Umbilical
Cord
LX-150
Light Source
Adaptor
65015
Endoscope
Connector Port