beautypg.com

2 contour analysis, 1 contour parameters – Welch Allyn Means ECG Physicians Manual for CP Series Electrocardiographs - User Manual User Manual

Page 8

background image

MEANS Physicians Manual


Welch Allyn

8

2 Contour analysis

2.1 Contour parameters

All parameters that are used in the diagnostic criteria of the contour classification are
measured in the representative P-QRS-T complex. The lead-independent, overall parameters
are presented in Table 1.

Table 1. Lead-independent parameters for the contour classification.

Name

Description

Heart rate

Ventricular rate (in beats per minute, BPM)

P axis

Axis of the P wave (in degrees, from

–180 to 180)

P duration

Duration of the P wave (in ms)

PR interval

Duration of the PR interval (in ms)

QRS axis

Axis of the QRS complex (in degrees, from

180 to 180)

QRS duration

Duration of the QRS complex (in ms)

Corrected QT interval

QT interval corrected for heart rate according to Bazett’s formula:
QTc = QT *

(HR/60) (in ms)


Hodges’ formula:
QTc = QT + 1.75 × (HR-60)

Note: The CP 50, CP 100 and CP 200, and the CP 150 and
CP 250 devices support either the Bazett or Hodges QTc
calculation on the printout. MEANS always uses the Bazett
calculation in its interpretive output statements.