4 diagnostic classification, 5 outline of the manual – Welch Allyn Means ECG Physicians Manual for CP Series Electrocardiographs - User Manual User Manual
Page 6

MEANS Physicians Manual 
 
 
Welch Allyn 
6
1.4 Diagnostic classification
The diagnostic logic operates on the parameters and produces both a rhythm classification 
and a contour or morphology classification. The criteria used by the computer may differ from 
the criteria used in the ECG textbooks. The basic reason is that a human observer is 
inaccurate but flexible and creative, a computer precise and obedient but rigid in its operation. 
There are several specific reasons why ECG criteria in the program may differ from the 
conventional ones. First, there is no uniformity of criteria in the literature. Then, criteria may 
be based on inaccurate measurement by eye. Also, ECG measurements may be “falsified” for 
the ease of the reader: axis calculations are generally made from the amplitudes of QRS 
complexes rather than from the surface areas under the QRS tracings as prescribed by 
theory. Further, criteria are sometimes not quantitatively defined (How flat must a flat ST-T 
be? How slurred is a slurred QRS upstroke?) or their measurement is not unequivocally 
prescribed. For the computer program to work, a quantitative definition must somehow be 
decided upon. Moreover, conventional criteria may have been based on measurements 
produced by technically outdated instrumentation. The amplitudes of R waves have been 
consistently underestimated, especially in children, due to filtering effects by too low 
frequency response of the electrocardiographs. Finally, a human interpreter may deviate from 
strict criteria as he sees fit: sometimes criteria have been made to meet a priori expectations. 
In one respect, the computer is inferior to the human observer: although the computer can 
measure very accurately, its powers of pattern recognition are inferior. For instance, it will 
have great trouble in detecting a P wave buried in a ST segment which is easily seen by the 
human eye. 
1.5 Outline of the manual
The following of this manual consists of two main parts. One part describes the diagnostic 
criteria that are employed in the contour classification of the Modular ECG Analysis System 
(MEANS), the other describes the criteria used in the rhythm classification of MEANS. Each 
part contains a brief introductory section, a description of the measurements that are used in 
the diagnostic logic, and a comprehensive list of statements and corresponding diagnostic 
criteria. Related statements have been grouped in sections, e.g., all statements related to 
intraventricular conduction delay, left ventricular hypertrophy, etc. Finally, an index of the 
statements that can be generated by the program is provided on page 55. 
A general format is used to specify the diagnostic criteria. The statement is given first, 
followed by one or more conditions that must be fulfilled for the statement to be issued by the 
program. Multiple conditions are combined with the use of logical “and” and “or” connectives, 
binding the (combinations of) conditions that have the same level of indentation. For example: 
Say:
“probable inferior infarct”
if:
Q duration
40 ms and 0.2
Q/R ratio < 0.3 in aVF
or
30
Q duration < 40 ms and Q/R ratio
0.3 in aVF
or
Q duration in aVF
20 ms
and
Q duration
50 ms and Q amplitude > 300 µV in III
