beautypg.com

Ref 73305, 73322, 73324, 73326 (power supplies), Outlet type, Instructions for use – Welch Allyn 6V Power Supply, Rectal Light Handle - User Manual User Manual

Page 6

background image

1-3

Instructions for Use

The Rectal Light Handle’s power supply (REF 73305, 73322, 73324, 73326)
provides power from an appropriate wall outlet to the Rectal Light Handle
(REF 73211) which houses the lamp (REF 07800). The Rectal Light Handle and its
appropriate power supply provide an illumination system that is intended for use
with Welch Allyn devices and/or accessories (see table on page 1-6).

REF 73305, 73322, 73324, 73326 (Power Supplies)

Outlet Type

Insert the Power Supply into the receptacle of
an appropriate voltage and plug configuration.

Once properly connected, the green LED
located on the power supply housing will
illuminate indicating power is present to the
power supply cord.

Verify that the power supply will connect to the appropriate outlet and that the
power supplied to the outlet is within the range specified in the table provided.

REF

Nom. Input

Voltage

Input Current

at Nom. Input

Voltage (mA)

Input Voltage

Range (V.)

Plug Style

73305

120V 60 Hz

71

110 - 130

A

73322

230V 50 Hz

30

210 - 230

B

73324

240V 50 Hz

27

230 - 250

C

73326

234V 50 Hz

27

230 - 250

D

A

B

C

D

USA

NEMA 1-15P

EUROPE

CEE 7/16

UK

BS 1363

AUSTRALIA /
NEW ZEALAND

AS 3112