1-9
3.
Follow one of the reprocessing methods described below.
4.
After performing a reprocessing procedure, wipe the lamp with a soft, dry,
lint free cloth.
5.
Re-attach the clean sleeve to the handle by screwing in a clockwise direction.
6.
Re-connect the power supply cord to the light handle and verify that the unit
works by plugging it in. If the lamp fails to illuminate follow the
troubleshooting instructions located in light handle manual (
700791)
supplied with your handle.
Cleaning (73210, 73211 and 73500)
1.
Prepare an enzymatic detergent formulated for endoscopic instru-
ments according to the manufacturer’s instructions.These instruc-
tions were validated using ENZOL/CIDEZYME.
2.
Submerge the item in the cleaning solution.
3.
Using a soft bristle brush, scrub the item submerged in the cleaning solution
for a minimum of 5 minutes then rinse with sterile water for 5 minutes.
Intermediate Level Disinfection (73210, 73211 and 73500)
1.
Follow the Cleaning instructions above.
2.
Using CaviWipes
™
, wipe all item surfaces and keep wet with CaviWipe
solution for a minimum of 5 minutes.
3.
Allow the item surface to dry for a minimum of 10 minutes.
Sterilization (73211 ONLY)
Caution Autoclaving the light handle may reduce the product’s
functional life.
CW
CCW
Rectal Light Handle Outer
Sleeve
Rectal Light Handle Body and
Cord