Welch Allyn Fiber optic laryngoscope handles - User Manual User Manual
Page 3
Directions for use
Welch Allyn fiber optic laryngoscope handles
3
Cleaning and
intermediate level
disinfection:
Drying:
Maintenance,
Inspection and Testing
1. Follow the germicidal wipe manufacturer’s instructions
to clean all exposed surfaces of the main handle, end
cap, and lamp cartridge.
2. If necessary, brush with a dry, soft-bristled brush and
re-wipe to loosen/remove excessive visible soil.
3. After all visible soil is removed, re-wipe to wet all
surfaces and allow adequate contact time for
disinfection as directed by the germicidal wipe
manufacturer.
CAUTION: Only use quaternary ammonium
isopropanol based germicidal wipes.
Allow components to air dry.
1. Inspect each component area (per figure 3) for damage or
deterioration.
WARNING: Discard any component that shows
evidence of damage or deterioration.
Contact Welch Allyn for component replacement.
2. Reassemble handle (per figure 4) with new or
batteries in known good condition.
3 . Attach handle to a clean and disinfected test blade
assembly in known working condition. Verify that:
•
Blade assembly engages and locks onto handle.
•
Blade assembly deploys into its locked position on
handle AND lamp illuminates.
•
Light output is satisfactory.
If the lamp fails to light or output is low, check or
replace the lamp and/or batteries.
Storage
Store handle per facility practice to allow device to remain
clean, dry, and ready for service.
End of reprocessing instructions for intermediate level disinfection.