Directions for use

Welch Allyn fiber optic laryngoscope handles

3


Cleaning and
intermediate level
disinfection:








Drying:


Maintenance,
Inspection and Testing


1. Follow the germicidal wipe manufacturer’s instructions

to clean all exposed surfaces of the main handle, end
cap, and lamp cartridge.

2. If necessary, brush with a dry, soft-bristled brush and

re-wipe to loosen/remove excessive visible soil.

3. After all visible soil is removed, re-wipe to wet all

surfaces and allow adequate contact time for
disinfection as directed by the germicidal wipe
manufacturer.

CAUTION: Only use quaternary ammonium

isopropanol based germicidal wipes.

Allow components to air dry.


1. Inspect each component area (per figure 3) for damage or

deterioration.

WARNING: Discard any component that shows

evidence of damage or deterioration.

Contact Welch Allyn for component replacement.

2. Reassemble handle (per figure 4) with new or

batteries in known good condition.

3 . Attach handle to a clean and disinfected test blade

assembly in known working condition. Verify that:

Blade assembly engages and locks onto handle.

Blade assembly deploys into its locked position on
handle AND lamp illuminates.

Light output is satisfactory.

If the lamp fails to light or output is low, check or
replace the lamp and/or batteries.

Storage

Store handle per facility practice to allow device to remain
clean, dry, and ready for service.

End of reprocessing instructions for intermediate level disinfection.