Directions for use

Welch Allyn fiber optic laryngoscope handles


Reprocessing instructions

These reprocessing instructions refer to procedures for cleaning and intermediate level
disinfection. Fiber optic laryngoscope handles must be reprocessed prior to first use and
between each use using the following method as outlined in this document:

Cleaning and intermediate level disinfection

Welch Allyn has validated the above instruction as being capable of preparing these
laryngoscope handles for re-use. The user must ensure that the reprocessing as actually
performed by the user's personnel, with the user's equipment and materials, achieves the
desired result. This may require validation and routine monitoring of the user's actual process

NOTE: The main handle and bottom cap components are compatible with the disinfection
solution and autoclave methods identified which are provided for facilities who wish to perform
either method after cleaning and intermediate level disinfection.

Cleaning and intermediate level disinfection instructions

Point of use: 1. Separate blade assembly from handle and place handle

into suitable containment for subsequent reprocessing
(per figure 1). Do not place handle with sharp devices.

2. Prevent the handle from drying (i.e. wrap/cover in moist

germicidal wipe).

Preparation for
decontamination:





























1. Select an appropriate quaternary ammonium

isopropanol based germicidal cleaner labeled suitable
for use on healthcare equipment and capable
intermediate level disinfection. Reference EPA-
registered disinfectants:
http://www.epa.gov/oppad001/chemregindex.htm
Outside of the U.S., please consult applicable
regulatory body for equivalent quaternary ammonium
isopropanol germicidal cleaner.

2. Remove batteries and lamp cartridge (per figure 2).