Directions for use

Welch Allyn rechargeable fiber optic and standard

5

laryngoscope

handles

Fiber Optic handles, 60713 (2.5v) & 60835 (3.5v)

CAUTION: Use only with Welch Allyn REF: 7114X Universal Charger.

1. Set up and operate the REF: 7114X Universal Charger per its instructions.
2. After initial cleaning and disinfection, place assembled handle(s) in charging well(s). See Figure 5.

The charger’s yellow indicator LED illuminates, indicating that the handle is charging. The
indicator LED is located directly under the handle well.

3. Charge handle for approximately 16 hours or overnight to assure battery is fully charged.
4. After charging, continue maintenance, inspection, testing and intermediate level disinfection per

these instructions.

Notes

 Slight heating of handles is normal during charge cycle.

 Batteries cannot overheat or overcharge.

 Batteries will not leak.

 Batteries will discharge gradually over a period of about 60 days when stored at room temperature

(70

o

F). Storage at higher temperatures accelerates the discharge rate.