Reprocessing instructions – Welch Allyn Rechargeable laryngoscope handles - User Manual User Manual
Page 2
![background image](/manuals/641577/2/background.png)
2
Directions for use
Welch Allyn rechargeable fiber optic and standard
laryngoscope
handles
Reprocessing instructions
These reprocessing instructions refer to procedures for cleaning and intermediate level disinfection.
Rechargeable laryngoscope handles must be reprocessed prior to first use and between each use
using the following method as outlined in this document:
•
Cleaning and intermediate level disinfection
Welch Allyn has validated the above instruction as being capable of preparing these laryngoscope
handles for re-use. The user must ensure that the reprocessing as actually performed by the user's
personnel, with the user's equipment and materials, achieves the desired result. This may require
validation and routine monitoring of the user's actual process
Cleaning and intermediate level disinfection instructions
Point of use:
1. Separate blade assembly from handle (see Figure 1)
and place handle into suitable containment for
subsequent reprocessing. Do not place handle with
sharp devices.
2. Prevent the handle from drying (i.e. wrap/cover in moist
germicidal wipe).
Preparation for
decontamination:
1. Select an appropriate quaternary ammonium
isopropanol-based germicidal cleaning wipe labeled
suitable for use on healthcare equipment and capable
of intermediate level disinfection. Reference EPA-
registered disinfectants:
http://www.epa.gov/oppad001/chemregindex.htm
Outside of the U.S., please consult applicable
regulatory body for equivalent quaternary ammonium
isopropanol germicidal cleaner.
2. For 60710 & 60720 standard handle; unscrew bottom
cap (5) of handle counterclockwise, remove battery
(4) and separate handle (2 and 3). See Figure 2.
For 60713 & 60835 fiber optic handles; unscrew
bottom cap (5) of handle counterclockwise, remove
battery (4) and lamp cartridge (3). See Figure 3.
3. Set battery aside.
Figure 2
Figure 1
Figure 3