Rs -1 baud – Torbal BAS Series User Manual
Page 35
- 34 -
Example:
To set 1000g in weight B1.5 (d=0.5g) type:
S L 1 0 0 0 . 0 CR LF (53h 4Ch 31h 30h 30h 30h 2Eh 30h 0Dh 0Ah).
To set 100kg in weight B150 (d=50g) type:
S L 1 0 0 . 0 0 CR LF (53h 4Ch 31h 30h 30h 2Eh 30h 30h 0Dh 0Ah),
Setting the threshold 2 (optional):
Computer
→Scale: S H D1...DN CR LF (53h 48h D1...DN 0Dh 0Ah),
where: D1...DN – threshold value, up to 8 characters,
Scale
→Computer: no response
Cable WK-1 Configuration
Cable WD-1 Configuration
17.2 Baud Rate
1. Press the MENU key, when the option “Port - 1” is displayed press the T key, and then select the option
“Baud” by press the T key once again.
2. The scale will begin to display available baud rates: “4800”, “9600”, “19200”, “38400”, and “57600”. To
make a selection, press the T key when the desired setting is displayed.
rs -1
bAUd
4800
9600
19200
38400
57600