About this manual, 0 introduction, Introduction – Rice Lake Tracer AVi - v1.0 User Manual
Page 5

Introduction
1
About This Manual
This manual is intended for use by service technicians 
responsible for installing and servicing the 
Tracer AVi
single channel diagnostic junction box and bagwell 
display. 
Authorized distributors and their employees 
can view or download this manual from the 
Rice Lake Weighing Systems distributor 
site at 
www.ricelake.com
.
Some procedures described in this manual 
require work inside the 
iQube
enclosure.
These procedures are to be performed by 
qualified service personnel only.
The
Tracer AVi
does not have an on/off switch and therefore
must be installed near a power outlet that is easily accessible 
and in accordance with UL/CSA Safety Standards. 
1.0
Introduction
The
Tracer AVi
system consists of three components:
•
iQube
TM
, a digital programmable junction box
used with an analog load cell. It outputs a 
serial string that can be directly input to the 
remote display and 
VIRTUi
.
•
RD-1 remote display, a .8”, six-digit, 
seven-segment LED display.
•
VIRTUi
TM
, a PC-based indicator for
iQube
.
The virtual front panel consists of display and 
five-button keypad.
The
iQube
consists of three boards:
•
The connector board provides the physical 
c o n n e c t i o n s f o r t h e l o a d c e l l , s e r i a l 
communications, and power. 
•
The core module, which plugs into the 
connector board, contains the 
Tracer AVi
’s
processor and stores configuration and 
calibration data for the 
Tracer AVi
. The core
module provides a discrete A/D input and 
converts the analog load cell signal to a digital 
serial output.
•
The 7.5 VDC power supply mounts inside the 
enclosure and requires a 115/230 VAC input.
Configuration
The
Tracer AVi
system can be configured by using a
PC running the
VIRTUi
configuration program. This
method defines the load cell connected to
iQube
,
which is associated with a platform, and the platform 
t h a t m a k e s u p t h e s c a l e . E v e n t h o u g h t h e 
single-channel 
iQube
is just one cell, one platform,
and one scale; the association of scale and platform is 
required because the software is also used on multiple 
cell and multiple platform scales. 
 
Configuration consists of the following steps:
Define Load Cell: This is the electrical sensitivity 
(mV/V output) and capacity specification of the load 
cell. Load cell name and serial number can also be 
specified. 
Define Platforms: The
iQube
board must assign a
load cell to the platform.
Define System: The
iQube
board must assign the
platform to the system.
Load Cell Trimming and Calibration
iQube
supports calibration of multiple load cells with
corner match and section match calibration.
For the single cell
iQube
, only multi-point linear
calibration and theoretical calibration apply. Each type 
of calibration captures the initial dead load of the scale 
and provides a means to trim the output of the load 
cell.
Based on the cell capacity and sensitivity, the 
theoretical calibration calculates weight values based 
on the total signal from the cell.
Single Cell Diagnostic
Diagnostic functionality can be enabled for the
iQube
to identify abnormal load cell output. The diagnostic 
conditions that can be identified are open bridge open 
channel, drifting, peak-to-peak noise, cell at rail, and 
cell over/under range.
Error conditions generate a displayed error message if 
connected to 
VIRTUi
which can be configured to
email the alert message to an address.
