Command format – Rice Lake TP Series, Rice Lake Weighing Systems - Software Development Kit User Manual
Page 16

12 TP Series
Command Format
1. Composed of four characters (ASCII), including terminators (CR=0DH,
LF=OAH)
2. Instruction for tare subtraction (zero adjustment)
3. Setup of output control
* The output controls executed with commands [O0-O7] work the same as the output
controls executed through function setup on the main unit of the balance. The
commands [O8] and [O9] are data request commands issued to the balance.
** Once any command from [O0] to [O9] is exectued, the balance runs that function
until another command is entered. However, if the balance is switched off and on
again, the output control is reset to the initial function (function set value).
1
2
3
4
C1
C2
CR
LF
C1
C2
Code
Description
Value
Response
T
(SP) 54H
20H Instruction for
tare subtraction
(zero adjustment)
None
A00: Normal end
E01: Tare subtraction cannot
be executed due to an error in
the weight value
C1
C2
Code
Description
O
0
4FH
30H Stop output.
O
1
4FH
31H Output continuous at all times.
O
2
4FH
32H Output continuous if stable (stops output if unstable).
O
3
4FH
33H Outputs once by pressing Memory (irrespective of whether
stable).
O
4
4FH
34H Outputs once, if stable. Outputs if the balance is stable
when a sample is loaded after the preceding sample has
been removed and the balance indicated zero, or less.
O
5
4FH
35H Outputs once if stable, and stops output when unstable.
Even if the sample is not replaced, the balance is output
once it stabilizes the next time (including the zero indication).
O
6
4FH
36H Outputs once if stable, and outputs continuously when
unstable. Even if the sample i snot replaced, output of the
balance stops when it stabilizes after being output once.
O
7
4FH
37H Pressing Memory key causes the balance to output once
when stable.
O
8
4FH
38H Output once immediately.
O
9
4FH
39H Output once after stabilization.