Maintenance and replacement parts, Load cell wiring – Rice Lake RL1800 Series Stainless Steel User Manual
Page 7
7
3. Load Cell Wiring
1.
Route the load cell cables so they will not be damaged or cut. Cable should
not be routed near heat sources greater than 150
°
F (66
°
C). Do not shorten
any load cell cable. The load cell is temperature compensated with the
supplied length of cable. Cutting the cable will affect temperature compen-
sation. Coil excess cable and protect it so it will not be mechanically
damaged or be sitting in water.
2.
Provide a drip loop in all cables so that water or other liquids will not run
directly down the cables onto either
the load cells or the junction box.
Attach load cell cable to the dead
structure, not the vessel.
3.
If conduit protection is necessary
against mechanical or rodent damage
to the load cell cables, use flexible
conduit and conduit adapters at the
load cells.
4.
Connect cables for standard RL30000, RL39123, and RL35083 load cells
to the summing board in the junction box according to the guide shown
below and the labels on the terminal strips of the junction box. To verify
the wiring scheme, see the certification shipped with each load cell.
5.
For better performance, use positive and negative remote sense lines if the
wiring run from the junction box to the indicator is longer than 25 feet.
LOAD CELL WIRE COLOR
FUNCTION
Black
+EXC
Red
- EXC
Green
+SIG
White
- SIG
SHIELD
Gray or Bare
DRIP LOOP
4. Junction Box Connections, Adjustments & Calibration
•
Refer to junction box manual for trimming details.
•
Refer to indicator manual for system calibration guidelines.