5 peak hold, 1 capture peak force – Rice Lake MSI-8000HD Remote Display User Manual
Page 17

User Defined Function Keys 13
3.5 Peak Hold
Peak Hold automatically updates the display when a higher peak weight reading is
established.
The Peak Hold function uses a high speed mode of the A/D converter (220 samples)
allowing it to capture transient loads at a far higher rate than typical Dynamometers.
Peak hold is cleared and re-enabled with the
Fx-Peak Hold
Key. When a new peak is
detected, the Fx LED will flash three times. The accuracy of the system in Peak Hold
mode is slightly reduced to .2% of Capacity + 5d. The Filter setting is turned off while
in Peak Hold mode to ensure the fastest acquisition rate.
Example
The Peak Hold function is useful in “Dynamic” and “Fall” tests. Common tests
include “Overall Breaking Strain” (OB€), Breaking Force, and Cycled Breaking
Strain.
3.5.1
Capture Peak Force
1)
Program a function key to Peak
Hold (P-Hld)
In this example we’ll use
F1
for Peak
Hold.
2)
Prepare the test stand and test
sample.
3)
Press
ZERO
to zero out any
residual strain on the link.
4)
Press
F1 PkHold
Confirm that
the “Pk” annunciator is on.
A small jump in the reading may occur
depending on the stability of your test
device.
5)
Apply the test force. The F1 LED
will blink three times when a
new peak value ID is detected.
6)
When the force is removed,
the Peak value can be recorded.
7)
To run a new test, press
F1 PkHold
to clear the Peak
Value. Confirm the “Pk”
annunciator is off. Then repeat
steps 3-6.
Peak Hold
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
Pk
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
Pk
Example Peak Captured Value
Peak Hold
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
kg
lb
M
kN
Ton
Peak Cleared Value
F1
F1
0
ZERO
Figure 0-1. Capture Peak Force