9.0
Troubleshooting & Testing
Refer to the following instructions to check and correct any failure before contacting service personnel.
Table 9-1. Troubleshooting Table for the Rice Lake Scale Line
Symptom
Possible Cause
Corrective Action
Scale does not turn on
Dead battery
Connect the scale to a power source.
Faulty electrical outlet
Use a different electrical outlet.
Bad power supply
Replace adaptor.
Questionable weight or the scale
does not zero
External object is interfering with the
scale
Remove the interfering object from the scale.
Display did not show 0.0 before
weighing
Zero the scale and begin the weighing process
again.
Scale is not placed on a level floor
Ensure the scale is level and begin the weighing
process again.
Scale is out of calibration
Check the weight with a known weight value.
Improper tare
Place the tare item on the scale. Once the weight
of the item is displayed, press TARE.
The display shows a STOP message
The load on the scale exceeds the
capacity of the scale
Remove the excess weight and use the scale
according to manufacture’s specs.
The display shows LO Bat message
The battery is low
Recharge the battery.
The display shows Err message as
detailed below
Err 2
Low saturation state (low A/D)
The load cell is not connected properly. Check
the cables and mechanical connections. If the
problem persists, replace the set of load cells.
Err 3
High saturation state (high A/D)
See Err 2
Err 6
Unstable weight. Cannot calibrate
Check the load cells’ mechanical surroundings
and see that nothing touches them and that the
cables are properly welded.
Err 7
Mathematical error; division by zero.
Cannot calculate calibration factor.
This error will show when trying to calibrate the
unit with no calibration weight on the unit.
Calibration or configuration settings
are messed up
Pushed the wrong keys
Reset to factory defaults. See Page 9 for
instructions.
The display shows a Con En message The configuration jumper is not in
place.
Install the configuration jumper on both pins. (see
Figure 4-1 on page 7 for more information).
9.1
Test Mode
The test mode menu is a special mode used for checking three very important parameters which are useful in
knowing the system’s state and for troubleshooting. The configuration/calibration jumper must be removed
before you can enter this mode (see Figure 4-1) and power must be off.
To access the test mode parameter, press and hold the
REWEIGH
+
On-Off/Zero
keys simultaneously until middle
arrows appear in the display.
The test mode has three parameters. They are:
•
Weighing
•
Internal count
•
Battery indication.
Alternating between the parameters is performed by pressing the
REWEIGH
key.
Press the
Kg-Lb
key to zero the scale while displaying weight (
TEST
).