Rice Lake Discovery Series Semi-Micro and Analytical, Ohaus User Manual
Page 39

EN-35
Ohaus Discovery
3.7.3 Setup
SETUP
Legal Trade On/
Off
Auto Tare On/
Off
Auto Off
Off, 1, 2, 5 min
Back Light
On/Off
End Setup
Legal Trade
When set On, the balance operates in compliance with Weights and
Measures regulations.
When the LEGAL TRADE menu item setting is set to ON, the following
conditions apply:
• The CALIBRATION menu is hidden.
• The LEGAL TRADE menu item is hidden.
• The AUTO ZERO menu item setting is set to 0.5d and locked.
• The OUTPUT WHEN STABLE menu item setting is set to ON
and locked.
• The CONTINUOUS setting in the AUTO PRINT menu item is
hidden.
Auto Tare
When set On, the balance will automatically tare the first item placed
on the pan.
Auto Off
When set to 1, 2 or 5 minutes, the balance will turn off in the
selected time interval if there has been no activity. Set to OFF for
continuous operation.
Backlight
When set On, display is illuminated at all times with balance on.
End Setup
Press the
Yes button, to advance to the Readout submenu. Press
the
No button to return to the LEGAL TRADE menu item.
3.7.4 Readout
Auto Zero
Allows setting the balance auto zero level setting: Off, .5, 1, 2 or 5
divisions. Balance maintains zero until threshold is exceeded.
Filter
Allows setting the balance filter level: Low, Medium, High. Compen-
sates for vibration or excessive air currents.
READOUT
Auto Zero Off, .
5d, 1d, 2d, 5d
Filter Low,
Medium, High
Gross Ind. G, B,
Off
End Read
3.7.5 Mode
Gross Ind.
Allows setting the gross indicator to G (gross), B (brutto) or Off.
End Read
Press the
Yes button to advance to the Mode sub-menu. Press No to
return to Auto Zero.
MODE
Weigh
On/Off
Count On/
Off
Apw Optimize On/Off
End Count
Percent On/
Off
Check Weigh On/
Off
Animal On/
Off
Filter
Low, medium, High
End Animal
GNT On/
Off
Totalize On/
Off
High Point On/
Off
Density On/
Off
D Mode
Liquid, Solid
Aux Liquid
Water, Other
Porous
On/Off
End Density
Statistics On/
Off
S Mode
Weigh, Check,
Animal
Low, Medium, High
Pipette On/
Off
Set Liquid
Water, Other
Set B Unit
Atm, Inhg, Hpa,
Mbar, Mmhg, Psia
Set N Unit
ML, UL
End Pipette
End Mode
The Mode sub-menu is used to turn Application modes On or Off.
Weigh
Use to set the Weighing Mode On or Off.
80104130.indd 35
11/7/2006 9:22:44 AM