SB-1150 Installation & Operation Manual - Maintenance
27
11.3
Load Cell Replacement
Replacement load cells can be ordered from Rice Lake Weighing Systems. To replace a load cell, lift the scale
with a chain and remove foot, then remove the defective load cell. Disconnect load cell cable from the junction
box and cut cable ties. When the cable is freed, pull cable out of the scale frame channels.
Figure 11-1. Load Cell Assembly
Follow the directions given below to install new load cells.
Lay out the four load cells near the corners where they are to be installed. Thread the cable from each load cell
through the conduit tubing in the frame and into the junction box according to the wiring diagram in Figure 9-3.
Note that in Figure 11-2 both the scale and the junction box are viewed from the bottom. To verify correct load
cell/junction box terminal matching, see the numbers on the terminals inside the junction box and the corner
numbering diagram in Figure 11-4 28.
Figure 11-2. Bottom View of Scale
Check that the threaded holes for the load cell screws are free of debris. Use compressed air to blow out holes if
necessary. Position load cells with alignment arrows pointed up toward the deck and loosely install the hex head
cap screws provided, as shown in Figure 11-1. If the base is used with a pit frame or access ramp, position the
load cell to maintain the dimension shown in Figure 11-3. With the torque wrench, tighten all bolts to 75 ft-lb.
CELL 4
CELL 3
CELL 2
CELL 1
1
4
2
3