0 test menu, Test menu – Rice Lake RL550 Digital Chair Scale Technical and Operating Instructions User Manual
Page 22

18
Rice Lake 550 Digital Chair Scale Technical Manual
9. Once all the digits have been entered, press the
ENTER
key and
Clear
is displayed.
10. Make sure the scale platform is clear of weight and press the
ENTER
key again and ===== is displayed.
11. A request to put the chosen load on the platform is displayed by
Put xxx.xx
.
12. Put the chosen weight on the platform and press the
ENTER
key. ===== is displayed and then
Save
.
13. Press the
ENTER
again and the display indicates
Done
.
14. Press the
BMI
key three times to exit back out to the top level
Done
parameter.
15. Press the
ENTER
key to return to weigh mode.
To exit calibration without changing zero or span existing calibration, press the
CLEAR
key, then the
BMI
key.
6.0
Test Menu
To access the
TEST
menu, use the following steps.
1. Turn the scale on by pressing the
On/Off
key and the
ENTER
key simultaneously until
ID
flashes.
2. Press the
ENTER
key again.
3. Continue to press the
BMI
key to scroll through the various menu items.
4. Once complete, press the
BMI
key again and
Done
is displayed.
5. Press the
ENTER
key to start the weighing process.
DEF
CALIB
PROG
DONE
TEST
VER
BAT
VALUE
A2D
1.83
value
value
value
Figure 6-1. Test Menu
Table 6-1. Test Menu
Test Menu
Parameter
Choice
Description
VER
Value
Displays the current software version.
BAT
Value
Displays the current battery level.
VALUE
Value
Displays the actual value.
A2D
Value
Displays the actual raw counts of the scale.