Remote connections, Mounting the hi-6100 in a rack – ETS-Lindgren HI-6100 Field Monitor User Manual
Page 21

Assembly and Installation
|
21
Remote Connections
For remote-controlled operation of the HI-6100 connect the RS-232 or IEEE-488
cable to the appropriate connector.
If RS-232 remote control is used, see Remote Operation on page 47
for information on setting the device baud rate.
If IEEE-488 remote control is used, set the DEVICE ADDRESS to the
appropriate bus address for your installation. See Operation on
page 25 for information on setting the device address.
Test remote operation by issuing an *IDN? command from the remote controller
(*IDN?
Remote Operation on page 47 for additional information.
Mounting the HI-6100 in a Rack
This product is designed for use in Installation Category II and Pollution Degree II
per IEC-1010 and IEC-664. When installing the HI-6100 in a rack the convection
into and out of the product must not be restricted. The ambient temperature
(outside of the rack) must be less than the maximum operating temperature of
the HI-6100 by 4°C for every 100 Watts dissipated in the rack. If the total power
dissipated in the cabinet is greater than 800 Watts, then forced convection must
be used.