beautypg.com

Charging indicators – ETS-Lindgren HI-6153 Electric Field Probe User Manual

Page 88

background image

88

|

Series H-491198-01 Battery Charger for NiMH Batteries

C

HARGING

I

NDICATORS

The following LEDs are located on the front of the battery charger:

POWER ON (green)—Indicates the battery charger is connected to the
AC power source.

NO BATTERY (amber)—Indicates the battery charger does not detect
a battery.

PENDING (amber)—Indicates the battery charger detects a battery.

Before fast charging can begin, the battery voltage must fall within
predetermined acceptable limits. A pulse-trickle charge is provided to
bring a depleted battery to a valid charge prior to fast charge.

CHARGING (amber)—Indicates the voltage pre-qualification condition
has been met, and fast charge has started.

Fast charging continues until termination by a peak voltage detect or a
maximum charge time. For peak voltage detect, the fast charge is
terminated when the battery voltage is lower than the previously
measured values by 24mV. If a peak voltage is not detected, then the
fast charge is terminated when the maximum time limit of 180 minutes
is reached.

COMPLETE (green)—Indicates a fast charging peak voltage is
detected.

If the probe is turned on during the fast charge period of the charge
cycle, the current surge that the batteries provide to the electronics in
the probe will cause a momentary voltage drop on the batteries,
causing the charger to terminate the fast charge based on a peak
voltage detect condition.

The battery charger switches to top-off charge mode that provides a charge
current one-eighth of the fast charge rate. This charge continues from the time
when a peak voltage detect terminates the fast charge, until the maximum time of
180 minutes is reached. If the probe is disconnected from the charger at this
point in the charge cycle, the batteries will operate properly.