0 probe shield care and replacement – ETS-Lindgren HI-6153 Electric Field Probe User Manual
Page 81
Probe Shield Care and Replacement
|
81
7.0 Probe Shield Care and Replacement
If you have an HI-6122,
HI-6022, HI-6105, or
HI-6005 Field Probe, over
time the probe shields
(the cone-shaped
structures with axis
labels) may accumulate
foreign materials that
could become embedded
in the shield surface.
Probe shields must be
kept clean to maintain
proper operation.
If the probe shields need to be replaced, you may order them from
ETS-Lindgren. Follow these steps to replace the probe shields:
1. Remove the nylon screws from the probe shield.
2. Lift the shield straight up and away from the housing.
Lift the shield high enough to clear the internal antenna assembly, or
you may damage the assembly. Never touch the antenna assembly.
You may transfer a static charge to the assembly that may distort or
disable the measuring capability in that axis.
3. With the probe in a protective, supportive position, carefully replace the
probe shield by lowering it straight down onto the housing.
Make sure to clear the internal antenna assembly when lowering the
shield or you may damage the assembly.