Operating Protocols

Probe
Command

Description

Probe Response

D3

Read probe data

:Dx.xxxyy.yyzzz.zB

xxxx, yyyy, zzzz=

4-digit axis values with
floating decimal point

B=battery flag, N or F

D5

Read probe data

:Dx.xxxyy.yyzzz.zcccc.B

xxxx, yyyy, zzzz=

4-digit axis values with
floating decimal point

cccc=composite field value

with floating decimal point

B=battery flag, N or F

I

Identification command

:I60XXB

XX=specific probe model

sr=10-character software

revision

sn=8-character serial

number

cd=8-character calibration

date

B=battery flag, N or F

The Identification command,

I, may also be used as the first command

sent. The command will turn on the laser. Once communication
between the probe and the Laser Data Interface (LDI) is established,
the return string will be sent. Subsequent

I commands immediately

send the return string.

TC

Read temperature in
Centigrade

:Txxxx.