Page 102: Hi-61xx series laserpro field probes
102
|
Operating Protocols
I
NFORMATION
T
RANSFER
P
ROTOCOL
The probe responds to commands from another device; it transmits no data
without first receiving instructions to do so.
C
OMMAND
S
TRUCTURE
See the following pages for detailed information regarding the command
structure to the probe. When the probe completes the command, it responds with
a string consisting of:
A start character (":")
The command letter
Data (if required)
If the command does not require the probe to return any data, the probe simply
responds with the start character (":") then the command letter and a carriage
return. If an error occurs, the probe responds with an error code.
P
ROBE
C
OMMANDS
All probe commands return
:E7 when the probe is turned off.
HI-61XX
S
ERIES
L
ASER
P
RO
F
IELD
P
ROBES
Probe
Command
Description
Probe Response
B
Read probe converter
voltage
:Bxx.xx