Ets_initiatereadfield() – ETS-Lindgren HI-6153 Electric Field Probe User Manual
Page 132
132
|
ETSProbe DLL User Guide
ETS_I
NITIATE
R
EAD
F
IELD
()
int ETS_InitiateReadField(const int probeHandle);
Purpose:
Initiates a field reading from the probe. The function
returns immediately without waiting for the operation
finish.
To determine when the operation is finished, use
ETS_IsOperationComplete. When the reading is
completed use
ETS_Field or ETS_CombinedField to
obtain the field values.
The default field units are V/m or A/m.
Return Value:
Returns an integer status code. The numeric value
0
indicates successful completion.
See Status Codes on page 121 for error code
description.