beautypg.com

0 specifications – ETS-Lindgren HI-2790B Calibration Comparison System User Manual

Page 39

background image

www.ets-lindgren.com

Specifications

39

9.0 Specifications

Frequency:

2450 + 0.1 MHz over an ambient

temperature range of 20 degrees to

24 degrees C

Frequency Stability:

50 ppm/°C + 0.0025%

Dynamic Range:

0.1

– 5 + mW/cm

2

Plane-wave equivalent power

density at probe test location.

Input Power:

100

– 240 VAC

0.8

– 0.4 A max

50

– 60 Hz

Fuses:

250 Volt, 1.0 Amp, Type T

(5 mm x 20 mm)

Output:

2450 MHz, 4 W max

Do not operate without 50 ohm load.

Local Calibration Reference:

HI-1710 Microwave Survey System

(HI-1710 LCR)

Size:

Width: 770 mm wide

Depth: 1040 mm deep

Height: 1230 mm high

(1360 mm with HI-2790B

Power/Control Module)

Weight:

45 kg (shipping)

Support:

Castor wheels for ease of location