Calibration Comparison Log 30-Day Check

25

5.0 Calibration Comparison Log 30-Day Check

To make sure that microwave survey instruments are being checked in

accordance with Center for Devices and Radiological Health (CDRH) guidelines,

complete the Calibration Comparison Log 30 Day Check form on page 26 using

the following instructions, and in conjunction with the HI-2790B Calibration

Comparison System.

Note: The word MEAN is used to refer to the average or center

value of the polarization ellipticity reading.

5.1.1

A separate log should be kept for each instrument, identifying it by

model number and serial number.

5.1.2

Record the maximum, minimum, and mean values, and calculate

the ellipticity as follows:

Total % ellipticity = MAX minus MIN x 100

MEAN

5.1.3

Comparison readings should always be made at the mean and for

a given instrument. No reading may be lower than 10% below the

highest reading during the 12-month period between annual

calibrations.

The first reading in a 12-month period between annual

calibrations will always be acceptable if it is within the dB band

for all instruments.

The ellipticity must be within the manufacturer specifications.

For ETS-Lindgren instruments the limit is 10% total, plus or

minus 5%.

Compare the mean reading and the ellipticity reading against

these limits and determine acceptance or rejection.