Error messages – ETS-Lindgren HI-2790B Calibration Comparison System User Manual

Page 18

background image

18

Operation

www.ets-lindgren.com

Error Messages

Note: There are no user serviceable error messages.

The HI-2790B Power/Control Module continually monitors the power supply

voltages, the forward and reflected power levels, and the phase lock loop. If any

of these readings fall outside the correct operating range of the unit, the unit will

turn off the RF power and display one of the following messages:

10.5V LO

10.5V HI

6V LO

6V HI

5V LO

5V HI

REF HI

FOR HI

LOCK LOST RESET

If an error message appears, turn the HI-2790B Power/Control Module off to

reset the unit, and then turn it on again. If the error message persists, return the

unit to ETS-Lindgren for service.