CDRH Section 8 Report Guidelines

35

8.0 CDRH Section 8 Report Guidelines

Following is a suggested set of answers to Section 8 of the Center for Devices

and Radiological Health (CDRH) Guide for Preparing Reports on Radiation

Safety of Microwave Ovens.

8.1 Microwave Test Instruments (Type and Quantity)

To be answered by each oven manufacturer. List the manufacturer, model

number, and quantity of meters as requested.

For example: ETS-Lindgren Model HI-1501, quantity: 13

8.2 Daily Check

8.2.1

Check Yes.

8.2.2

Check Yes.

8.2.3

Check Yes.

8.2.4

The maximum allowable polarization ellipticity for all ETS-Lindgren

microwave survey meters is ± 5%, or 10% total, or 0.10 total. See

the manufacturer specifications for allowable ellipticity for other

models.

8.2.5

Check Yes.

8.2.6

Check Yes.

8.2.7

ETS-Lindgren Calibration Comparison System Model number

(insert model number) is used for performing the daily check on all

hand scan instruments. The daily check of the Automated

Microwave Oven Screening System (AMOSS) is performed in

accordance with the sign-on self-test criteria that are part of the

operational program of AMOSS.