Initial conditions – ETS-Lindgren HI-2790B Calibration Comparison System User Manual

Page 22

background image

22

Operation

www.ets-lindgren.com

Microwave Oven Survey Instrument Daily Ellipticity

Check

4.3.1

Zero the instrument to be checked, making sure the access door is

closed or the RF switch is off. Insert the probe of the instrument to

be measured into the HI-2790B probe holding fixture. Make sure

the probe is properly located against the stop on the holding

fixture. Set the scale of the instrument to that used for compliance

testing (usually the 0-2 mW/cm

2

range).

4.3.2

Turn RF POWER switch on and adjust the RF LEVEL control to

obtain a mid-scale reading (approximately 1 mW/cm

2

) on the

meter being checked.

4.3.3

Rotate the probe to obtain the minimum reading.

4.3.4

Readjust the RF POWER control so that the meter on the

instrument being checked indicates precisely 1.0 mW/cm

2

.

4.3.5

Rotate the probe to obtain the maximum reading.

4.3.6

If this value is less than 1.10 mW/cm

2

, the calibration status of the

meter is acceptable. If greater than 1.10 mW/cm

2

, it may be

rejected.

4.3.7

Record in a permanent log the daily checks on all instruments,

indicating whether each instrument was accepted or rejected.

Microwave Oven Survey Meter Calibration Comparison,

30-Day Check, Initial Conditions

4.4.1

Meters to be checked must be turned on and allowed to stabilize in

accordance with the manufacturer recommendations.

4.4.2

Select the desired probe range, zero the instrument, making sure

the access door is closed or the RF switch is off, and place the

instrument on top of the HI-2790B system to the left of the door.

4.4.3

Place the probe in the holder as described in step 4.1.1 on

page 19. Close the door.