Operation

23

Calibration Comparison

4.5.1

With the Power/Control Module RF POWER set to OFF, check

that the instrument being checked is properly zeroed. Turn the

RF Power switch on and check that the indicated RF POWER is

1.0 mW/cm

2

as established in step 4.1.9 on page 19. Rotate the

probe slowly through one full revolution (360 degrees) and record

the maximum and minimum readings.

4.5.2

Add the MAX reading to the MIN reading and divide the sum by

two. The result is the MEAN value.

4.5.3

Subtract the MIN reading from the MAX reading, divide the result

by the MEAN reading from the previous step, and then multiply

this result by 100. This is the TOTAL ELLIPTICITY in percent. The

maximum allowable ellipticity should be obtained from the

manufacturer specifications. Enter these values into the

Calibration Comparison Log.


Example: The test indicates a MAX value of 1.023 and a MIN value
of 0.985.

Add 1.023 and 0.985. This equals 2.008.

Divide 2.008 by 2. This gives a MEAN value of 1.004.

Subtract the MIN value (0.985) from the MAX value (1.023).

This gives a value of 0.038.

Divide 0.038 by the MEAN value (1.004), and then multiply the

result by 100. This gives a TOTAL ELLIPTICITY of 3.78%.