2.0 Maintenance

Before performing any maintenance, follow the safety

information in the ETS-Lindgren Product Information

Bulletin included with your shipment.

Regularly inspect all equipment and conduct scheduled

maintenance in accordance with the factory recommendations

provided.

WARRANTY

BEFORE SERVICING: CONTACT ETS-LINDGREN

(+1.512.531.6400)—Servicing or modifying the unit without

ETS-Lindgren authorization may void your warranty. If an

attempt to service the unit must be made, disconnect all

electrical power prior to beginning. Voltages exist at many points

within the instrument that could, if contacted, cause personal

injury. Only trained service personnel should perform

adjustments and/or service procedures upon this instrument.

Capacitors inside this instrument may still be charged even

when the instrument is disconnected from the power source.

Recommended Maintenance Schedule

Regular maintenance will prolong the effective operation and reliability of your

turntable. Follow the recommended schedule for monthly, 6-month, and

12-month service.

Do not perform maintenance while the turntable is operating.

M

ONTHLY

• Inspect the turntable assembly for loose fasteners, abnormal wear, and oil

leaks.

• Verify the accuracy of the current position setting on the ETS-Lindgren

controller in relation to the turntable. If the turntable always rotates in the

same direction, over time the 0-degree position may drift. To reset the

current position, see the controller manual.