ETS-Lindgren 2165 Turntable User Manual
Page 11

Maintenance | 11
2.0 Maintenance
Before performing any maintenance, follow the safety
information in the ETS-Lindgren Product Information
Bulletin included with your shipment.
Regularly inspect all equipment and conduct scheduled
maintenance in accordance with the factory recommendations
provided.
WARRANTY
BEFORE SERVICING: CONTACT ETS-LINDGREN
(+1.512.531.6400)—Servicing or modifying the unit without
ETS-Lindgren authorization may void your warranty. If an
attempt to service the unit must be made, disconnect all
electrical power prior to beginning. Voltages exist at many points
within the instrument that could, if contacted, cause personal
injury. Only trained service personnel should perform
adjustments and/or service procedures upon this instrument.
Capacitors inside this instrument may still be charged even
when the instrument is disconnected from the power source.
Recommended Maintenance Schedule
Regular maintenance will prolong the effective operation and reliability of your
turntable. Follow the recommended schedule for monthly, 6-month, and
12-month service.
Do not perform maintenance while the turntable is operating.
M
ONTHLY
• Inspect the turntable assembly for loose fasteners, abnormal wear, and oil
leaks.
• Verify the accuracy of the current position setting on the ETS-Lindgren
controller in relation to the turntable. If the turntable always rotates in the
same direction, over time the 0-degree position may drift. To reset the
current position, see the controller manual.