beautypg.com

Baumer BA UNxK09 IO User Manual

Page 8

background image

en_BA_UNXK 09-IO-Link.doc

8/18

Baumer Electric AG

24.04.2012/haa

Frauenfeld, Switzerland

Figure 2: Possible characteristic measuring value curves

5.1.3.1 Example numerical Teach-in:

1) The measuring range shall be defined from 50mm (A) to 180mm (B) -> (curve 2).

Point A absolute in tenth of mm:

500

 01F4 hex (= Measuring range limit A)

Point B absolute in tenth of mm:

1800

 0708 hex (= Measuring range limit B)

Parameter to be written:
Measuring range work:

01F40708 hex

2) The output signal shall be inverted; measuring range is from 60mm (B) to 120mm (A) -> (curve 3).

Point A absolute in tenth of mm:

1200

 04B0 hex (= Measuring range limit A)

Point B absolute in tenth of mm:

600

 0258 hex (= Measuring range limit B)

Parameter to be written:
Measuring range work:

04B00258 hex

30mm

200mm