Extended command check position „cx, 5 read / write offset „u“ (55h) – Baumer N 155 User Manual
Page 8
N 155
www.baumer.com
8
The following state-related values are available:
State
Significance
o (6Fh)
current value=target
x (78h)
current value≠ arget
Extended command Check Position „CX“
Besides the state value of the target/current value comparison, the extended command „Check Position“ reads
out the current value. The profile number is not transmitted. Regarding available state-related values please see
above.
Example:
SOH Adr Cmd
SCmd
EOT CRC
transmission
01h 20h 43h
58h 04h A8h
SOH Adr Cmd
state
reserved
reserved
Current value = -12.50
EOT CRC
Response
01h 20h 43h 6Fh 80h 80h 80h 80h 2Dh 30h 31h 32h 35h 30h 04h B7h
4.2.5 Read / write offset „U“ (55h)
This command is for reading or writing the offset to be added to the actual current value. This function however
has to be activated first in the command bit parameter ‚a’ (61h).
Example 1: Read offset
SOH
Adr
Cmd
EOT
CRC
Send
01h
20h
55h
04h
26h
Example 2: Write offset
SOH
Adr
Cmd
data (offset = –20.00)
EOT
CRC
Send
01h
20h
55h
2Dh
30h
32h
30h
30h
30h
04h
C3h
Response in both examples:
SOH
Adr
Cmd
data (offset = –20.00)
EOT
CRC
Response
01h
20h
55h
2Dh
30h
32h
30h
30h
30h
04h
C3h
4.2.6. Write a number sequence in upper line of the display „t“ (74h)
The command is for writing a 5-digit number sequence in the upper display line. The sequence is depicted with-
out dot respectively comma. Preceding zeroes as well as both arrows for direction are hidden. The bottom line is
still showing the current value.
The number sequence will remain until any command except „t“, „u“ or „R“ is received by interface.
Example: (number sequence = 054321)
SOH
Adr
Cmd
Number sequence = 054321
EOT
CRC
Send
01h
20h
74h
30h
35h
34h
33h
32h
31h
04h
C6h
SOH
Adr
Cmd
Number sequence = 054321
EOT
CRC
Response
01h
20h
74h
30h
35h
34h
33h
32h
31h
04h
C6h